Description:
- Microtitre wells are coated with the citrulinated synthetic peptides of 2. generation as an antigen.
- If specific antibodies are present, they bind to the antigen, are labeled by the Conjugate in the following steps and are detected by color reaction with a single component substrate (TMB-Complete).
- The kit allows 96 tests, including controls in a split microtiter plate with color-coded strips and breakable wells.
Advantages:
- The total assay time is about 2 hours.
- High sensitivity and specificity of the test.
- Kit includes CUT-OFF, Positive Control, Negative Control and Calibrators (10, 20, 130, 800 IU/ml)
- Semi-quantitative evaluation in the Index of Positivity (IP) or quantitative evaluation in IU/ml.
- Ready-to-use, color-coded components
- Single-component substrate
- Interchangeable components with the exception of kit specific components (Controls, Conjugate, Plate)
Application:
- Detection of high specific anti-CCP2 antibodies allows early diagnosis of Rheumatoid arthritis (RA).
- Checking of therapy results and determination of disease prognosis using the quantitative determination.
Brief assay procedure:
- Dilute samples (1:101)
- Pipette Controls and diluted Samples
- Incubate at 37°C for 30 min.
- Aspirate and wash the wells 5×
- Pipette Conjugate
- Incubate at 37°C for 30 min.
- Aspirate and wash the wells 5×
- Pipette Substrate (TMB-Complete)
- Incubate at 37°C for 30 min.
- Pipette Stop Solution
- Read color intensity at 450 nm
- Evaluate the test