Samples
· Capacity: up to 96 samples can be processed
· Dilution of patient samples in dilution strips
· Primary sample tubes: height 40-100 mm, diameter 12-16 mm
Robotic and pipetting system
· 1 pipetting arm - 2 types of tips (300 μl for samples/controls and 1300 μl for reagents)
· Possibility of two-stage sample dilution
· Tip trays: 108 for samples, 41 for reagents
· Pipetting volume: 25-1000 μl, volume control system
· Pipetting accuracy: < 3% CV (± 2% accuracy)
· Pipetting arm speed: samples < 19 minutes/plate (50 μl x 96 wells); two-stage dilution < 26 minutes; reagents < 2 minutes/plate
Washing
· 1 station
· 4 wash solutions with liquid-level monitoring
· Possibility of shaking (continuously or periodically)
· Waste: integrated container (8 l), level sensor
Measurement
· Spectrophotometer, 0-3,0 OD
· Spectral range: 405-690 nm
Incubation
· Up to 4 independent incubation units
· Temperature range: RT; 7 °C – 50 °C
· Stability: ± 1 °C
Software
· A part of PC delivery, OS Windows 7
· Bidirectional ASTM interface, XML, etc.
· Archiving patients results
· Quality control by Levey-Jennings, Westgard rules
Dimensions
· Depth: 91 cm; height: 80 cm; width: 106 cm
· Weight: mx. 136 kg
Certification
· CE marked, IVD certified